Opc - Agilent Technologies B2200A User Manual

Femto leakage switch mainframe; 14ch low leakage switch mainframe
Table of Contents

Advertisement

Syntax
Query response
Example
SCPI Command Reference

*OPC

*OPC
This command starts to monitor pending operations, and sets/clears the Operation
Complete (OPC) bit in the Standard Event Status Register as follows:
If there is no pending operation, sets the OPC bit to 1.
If there are any pending operations, sets the OPC bit to 0. The bit will be set to 1
again when all pending operations are completed.
So, *OPC command is required to enable the OPC bit. To stop monitoring pending
operations (disable OPC bit), execute the *CLS command.
For details about the Standard Event Status Register, see "Standard Event Status
Register" on page 5-58. Also, see *WAI command.
*OPC
1 <newline><^END>
*OPC? places ASCII character 1 into the Output Queue when all pending
operations are completed. For details, see "Output Queue" on page 5-60.
OUTPUT @Agb2200;"*OPC"
The following example is for query:
OUTPUT @Agb2200;"*OPC?"
ENTER @Agb2200;A
5-8
Agilent B2200 User's Guide, Edition 2