Connecting Input-Output Paths - Agilent Technologies B2200A User Manual

Femto leakage switch mainframe; 14ch low leakage switch mainframe
Table of Contents

Advertisement

Figure 4-2
Setup:
Programming
Programming Examples

Connecting Input-Output Paths

The following example connects instrument output to DUT as shown in Figure 4-2.
Input-Output Connection Example
Channel configuration mode: Normal
Connection rule: Single
Connection sequence: Break_Before_Make
Display strings: "Connecting MOSFET AG002201"
Used module: Switch module installed in the slot 1.
Connection paths:
from SMU1 to Output 1 (channel list 10101)
from SMU2 to Output 2 (channel list 10202)
from SMU3 to Output 3 (channel list 10303)
from SMU4 to Output 4 (channel list 10404)
4-8