Diagnostic:test:frame:state; [:Route]:Agnd:channel:disable:card - Agilent Technologies B2200A User Manual

Femto leakage switch mainframe; 14ch low leakage switch mainframe
Table of Contents

Advertisement

Syntax
Query response
Example
Syntax
Example
SCPI Command Reference

:DIAGnostic:TEST:FRAMe:STATe?

:DIAGnostic:TEST:FRAMe:STATe?
This command returns the most recent test result (pass/fail) of the specified test.
:DIAGnostic:TEST:FRAMe:STATe? CONTroller | FPANel | LED
| PEN | BEEPer
Parameter
CONTroller
FPANel
LED
PEN
BEEPer
test_result <newline><^END>
1: fail
0: pass
-1: not tested
OUTPUT @Agb2200;":DIAG:TEST:FRAM:STAT? CONT"
ENTER @Agb2200;A

[:ROUTe]:AGND:CHANnel:DISable:CARD

This command ground-disables the all output ports (channels) for the specified card.
When the Ground Mode is ON, the ground-disabled output ports are disconnected
from the input Ground Port. At *RST, no channel is ground-enabled. The ground
mode is set by "[:ROUTe]:AGND[:STATe]".
[:ROUTe]:AGND:CHANnel:DISable:CARD card_number
Parameter
card_number
OUTPUT @Agb2200;":ROUT:AGND:CHAN:DIS:CARD ALL"
5-24
Explanation
Controller test
Front panel interface test
LED matrix test
Light pen test
Beeper test
Explanation
For Auto Config mode: 0 or ALL
For Normal Config mode: 1, 2, 3, 4, or ALL
Agilent B2200 User's Guide, Edition 2

Advertisement

Table of Contents
loading

This manual is also suitable for:

B2201a

Table of Contents