Cls; Ese - Agilent Technologies B2200A User Manual

Femto leakage switch mainframe; 14ch low leakage switch mainframe
Table of Contents

Advertisement

Syntax
Example
Syntax
Query response
Semantics

*CLS

This command clears the Status Byte Register, the Standard Event Status Register,
and the Error Queue. This command does not clear the enable registers. See "Status
Reporting Structure" on page 5-53.
Also, this command stops the monitoring of pending operations by the *OPC
command.
This command does not have query form.
*CLS
OUTPUT @Agb2200;"*CLS"

*ESE

This command sets the bits of the Standard Event Status "Enable" Register.
1 enables, 0 masks.
*ESE enable_number
Parameter
enable_number
decimal integer (that is the sum of the binary-weighted values
for the desired bits), hexadecimal, octal, or binary value
enable_number <newline><^END>
enable_number is decimal integer value that is the sum of the binary-weighted
values of the "Enable" register bits.
The Standard Event Status "Enable" Register determines which bits of the Standard
Event Status Register are enabled. Enabled bits are ORed together, and the result is
reported to bit5 of the Status Byte Register.
The Standard Event Status "Enable" Register consists of 16 bits, but only the lower
8 bits are used, which correspond to the bits of the Standard Event Status Register.
For details, see "Standard Event Status Enable Register" on page 5-59.
Agilent B2200 User's Guide, Edition 2
SCPI Command Reference
Explanation
*CLS
5-5