Connection Sequence - Agilent Technologies B2200A User Manual

Femto leakage switch mainframe; 14ch low leakage switch mainframe
Table of Contents

Advertisement

NOTE
Table 3-1

Connection Sequence

This is only for modules that are set to the single connection rule. Refer to
"Connection Rule" on page 3-14.
Connection sequence specifies the open/close sequence of the relays when changing
from an existing connection to a new connection.
You can select one of the three connection sequences shown in Table 3-1 for each
module installed in the Agilent B2200.
When an existing connection is changed to a new connection, the relays are
opened/closed as follows:
Connection Sequence
Connection
Sequence
Break Before Make
Make Before Break
No Sequence
Agilent B2200 User's Guide, Edition 2
Relay Operation
1. Disconnect previous route.
2. Wait for relays to open.
3. Connect new route.
1. Connect new route.
2. Wait for relays to close.
3. Disconnect previous route.
1. Disconnect previous route.
2. Connect new route.
Front Panel Operation
Switch Control Functions
3-15