Introduction
Specifications
Supplemental Information for B2201A/B2211A
1
Offset current
2
IM noise (RMS)
Channel crosstalk capacitance
Offset voltage
3
Settling time
Bandwidth (at -3dB)
4
Guard capacitance
Additional C measurement error
1. When the voltage applied to all input-output paths is 0 V.
2. Measured by the Agilent 4156C with the integration time setting 100 PLC.
When the voltage applied to all other input-output paths is 0 V.
3. The time until the measurement value settles to within 300 fA of the final
value. When the applied voltage is 10 V.
4. When four modules have been installed in the mainframe. And when only
one input-output path is made.
5. For the measurement less than 1000 pF at the frequency 1 kHz to 1 MHz,
using the Agilent 4284A with 3 m cable. After the compensation using the
capacitance compensation algorithm (a function of VXIplug&play driver).
1-14
< 50 fA (SMU input)
5 fA (SMU input)
< 0.5 pF/channel (SMU input)
< 3 pF/channel (AUX input)
< 80 µV (SMU input)
< 100 µV (AUX input)
2.0 seconds
30 MHz (SMU input)
< 145 pF (SMU input)
5
< ± 1 % + 0.2 pF (SMU input)
Agilent B2200 User's Guide, Edition 2