To Create Compensation Data File - Agilent Technologies B2200A User Manual

Femto leakage switch mainframe; 14ch low leakage switch mainframe
Table of Contents

Advertisement

To Create Compensation Data File

This section explains how to create the compensation data file.
1. Select one of the compensation data files (template, 20 files) installed when the
Agilent B2200 VXIplug&play driver is installed.
To select the most appropriate template for your measurement environment, see
Table 4-9 that lists the file name and the measurement environment where the
template targets. Each template is a text file that contains the information as
shown below.
#
# Compensation data file for B2210A/Probecard I/F/3m triax cable
#
#
# MB
Mother Board
# MH
Matrix Path High
# ML
Matrix Path Low
# C1
Agilent Triax Cable
# C2H
Probe card I/F or User Triax cable High
# C2L
Probe card I/F or User Triax cable Low
# C3H
Probe card or User Coax cable High
# C3L
Probe card or User Coax cable Low
#
format version 1.0
B2210A
PCIF
#
#
R[Ohm]
#-----------------------------------------------------------
MB
0.000000e+00
MH
2.430000e+00
ML
2.490000e+00
C1
6.300000e-01
C2H
2.988000e-01
C2L
2.988000e-01
C3H
0.000000e+00
C3L
0.000000e+00
The 15th line specifies the used switch module, B2210A or B2211A.
The 16th line specifies the DUT interface, PCIF or CABLE. PCIF indicates that
the Agilent B2220A probe card interface is used. CABLE indicates that the
connector plate is used.
The lines C2H to C3L should be modified for each measurement environment.
See Table 4-9 and Table 4-10. Do not modify the other lines.
Agilent B2200 User's Guide, Edition 2
Capacitance Compensation
L[H]
C[F]
5.250000e-08
2.940000e-11
6.310000e-07
1.930000e-10
5.970000e-07
1.920000e-10
1.250000e-06
1.600000e-10
5.090000e-07
7.000000e-11
5.090000e-07
7.000000e-11
8.000000e-08
1.500000e-13
8.000000e-08
1.500000e-13
Programming
4-23

Advertisement

Table of Contents
loading

This manual is also suitable for:

B2201a

Table of Contents