Diagnostic:test:card:clear; Diagnostic:test:card[:Execute] - Agilent Technologies B2200A User Manual

Femto leakage switch mainframe; 14ch low leakage switch mainframe
Table of Contents

Advertisement

Syntax
Example
NOTE
Syntax
Query response
Example

:DIAGnostic:TEST:CARD:CLEar

This command clears the relay test result (pass/fail result) of the specified card.
:DIAGnostic:TEST:CARD:CLEar card_number
Parameter
card_number
OUTPUT @Agb2200;":DIAG:TEST:CARD:CLE 1"

:DIAGnostic:TEST:CARD[:EXECute]?

This command executes the relay test for specified card, then returns the pass/fail
result. Before starting the relay test, open the input/output terminals (end of cable is
OK, Kelvin cable must be removed). For more information, see "RELAY_TEST"
on page 3-35.
This command changes all settings of card to same state as after executing
:SYST:CPON command.
:DIAGnostic:TEST:CARD[:EXECute]? card_number
Parameter
card_number
test_result <newline><^END>
1: failure card exists
0: pass
OUTPUT @Agb2200;":DIAG:TEST:CARD? ALL"
Agilent B2200 User's Guide, Edition 2
:DIAGnostic:TEST:CARD:CLEar
Explanation
card number: 1, 2, 3, 4, or ALL
Explanation
card number: 1, 2, 3, 4, or ALL
SCPI Command Reference
5-21