Agilent Technologies B2200A User Manual page 84

Femto leakage switch mainframe; 14ch low leakage switch mainframe
Table of Contents

Advertisement

Table 3-5
Label
01 to 14
Shift
Remote
Lock
Error
Card#
Bias
Couple
Gnd
Rule
Sequence
NOTE
Front Panel Operation
Display Functions
LCD Display Items
Port function assigned to the input ports 1 to 14. B (bias port), C (couple port),
G (ground port), - (ground enabled port), or blank (no function).
Shift key status indicator. The triangle mark appears when the sub key is active.
Remote status indicator. The triangle mark appears when the Agilent B2200 is in
the GPIB remote condition.
Key status indicator. The triangle mark appears when the front panel keys are
locked by an external computer.
Error status indicator. The triangle mark appears when any error has occurred.
Slot number of the switch module to be monitored. Displays A when the channel
configuration mode is Auto. In the auto mode, the modules installed in the
mainframe from slot number 1 continuously will be controlled as one module. Or
displays 1 to 4 when the channel configuration mode is Normal. The module can
be selected by the
Card Selection
Bias mode status indicator. The triangle mark appears when the mode is ON.
Couple mode status indicator. The triangle mark appears when the mode is ON.
Ground mode status indicator. The triangle mark appears when the mode is ON.
Connection rule, FREE (plural route mode) or SINGLE (single route mode).
Connection sequence, BBM (Break Before Make), MBBR (Make Before Break), or
NO_SEQ (No Sequence).
In the remote mode and the default setting, only the Remote, Lock, and Error
indicators are available. To display all information, set the RMT_DSPL function
ON. Refer to "RMT_DSPL" on page 3-33.
3-24
Description
key.
Agilent B2200 User's Guide, Edition 2