Capacitance Compensation
To obtain compensation coefficients
Obtain the compensation coefficients as shown below.
1. Select the measurement frequency (Fmeas) used for the capacitance
measurement of a device under test (DUT), and set it to the Agilent 4284A. The
coefficients must be measured at the same frequency.
2. Perform the Agilent 4284A open calibration at the measurement terminal.
Optionally, perform short calibration if you want.
3. See Table 4-11 and Figure 4-7, and set the Agilent 4284A.
4. Connect the path/cable corresponding to C3H shown in Figure 4-6 to the Agilent
4284A. Then measure and record the R, L, and C values.
5. Connect the path/cable corresponding to C3L to the Agilent 4284A. Then
measure and record the R, L, and C values.
6. If you use the connector plate, perform the following procedure.
a. Connect the path/cable corresponding to C2H to the Agilent 4284A. Then
measure and record the R, L, and C values.
b. Connect the path/cable corresponding to C2L to the Agilent 4284A. Then
measure and record the R, L, and C values.
R, L, C Measurement Conditions
Parameter
R
L
C
a. Select 1 point. Do not change while measurements of all coefficients.
b. For triaxial cable, connect B to F directly, and measure L between A
and E. For coaxial cable, connect B to D directly, and measure L
between A and C. Ignore E and F.
Measurement Terminals of C2H/C2L/C3H/C3L Path
GUARD (C)
GROUND (E)
FORCE or SENSE (A)
4-26
Frequency
a
1 kHz to 1 MHz
Insulator
Triaxial Cable
Agilent B2200 User's Guide, Edition 2
Function
Terminals
A and B
SERIES
see note
PARALLEL
A and C
GROUND (F)
GUARD (D)
FORCE or SENSE (B)
b