Replacing Major Assemblies

A5 GPIB Handshake Data Input Assembly Removal/Replacement

A5 GPIB Handshake Data Input Assembly
Removal/Replacement
This module contains components that can be damaged or destroyed by
CAUTION
electrostatic discharge. It should be serviced at a static-safe
workstation. Refer to "Preparing a Static-Safe Work Station" in
Chapter 8.
To Remove the A5 GPIB Assembly
1. Remove the top cover
2. With a strong upward motion, lift the green ring at the top edge of
To Replace the A5 GPIB Assembly
1. Gently insert the board with a vertical motion into the socket
2. Replace the top cover
7-12 11713 Operating and Service Manual
the A5 GPIB handshake data input assembly.
marked A5 GPIB handshake data input assembly. Apply balanced
pressure with your thumbs to push the board into the socket.