Specifications
Transmission Rate Measurement Capabilities
Test Range, Package
Test Range, Film, 50 cm
Resolution, Package
Resolution, Film
Repeatability, Package
Repeatability, Film
Table 10.6: Transmission Rate Specifications - AQUATRAN 3/40 Model L
Note: A minimum Exam Time of 45 minutes is required to meet the specifications stated above.
10-4
0.000025 to 0.5 g/(pkg • day)
2
0.005 to 100 g/(m
0.000001 g/(pkg • day)
0.0002 g/( m
± 0.000025 g/(pkg • day) or 2% of value, whichever is greater
± 0.005 g/( m
Revision D
AQUATRAN Model 3/40 Operator's Manual
2
• day)
• day)
2
• day) or 2% of value, whichever is greater
2
MOCON, Inc.