Transmission Rate Measurement Capabilities - Ametek mocon AQUATRAN 3/40 Operator's Manual

Water vapor transmission rate system
Table of Contents

Advertisement

Specifications

Transmission Rate Measurement Capabilities

Test Range, Package
Test Range, Film, 50 cm
Resolution, Package
Resolution, Film
Repeatability, Package
Repeatability, Film
Table 10.6: Transmission Rate Specifications - AQUATRAN 3/40 Model L
Note: A minimum Exam Time of 45 minutes is required to meet the specifications stated above.
10-4
0.000025 to 0.5 g/(pkg • day)
2
0.005 to 100 g/(m
0.000001 g/(pkg • day)
0.0002 g/( m
± 0.000025 g/(pkg • day) or 2% of value, whichever is greater
± 0.005 g/( m
Revision D
AQUATRAN Model 3/40 Operator's Manual
2
• day)
• day)
2
• day) or 2% of value, whichever is greater
2
MOCON, Inc.