Maintenance
AQUATRAN Model 3/40 Operator's Manual

Maintaining the Test Cells

Periodically the test cell gas ports should be examined, and any excess buildup of grease should be
removed. Alcohol can be used to remove any residue from the grease. Inspect the gas ports and True Seal
flush port and remove any grease that could obstruct the flow of gas.
Periodically the test cell O-rings should be examined and any damaged, deformed, cracked or brittle O-
rings replaced. The parts of the test cell and the associated O-ring part numbers are illustrated in Figure 8-1
and
Table 8.1.

Test Station O-Rings and Part Numbers

The serviceable parts of the test station are shown in Figure 8-1 below. Refer to Table 8.1 for the
associated part numbers and the quantity used. To order replacement parts contact MOCON in the USA at
(763) 493-6370.
2
1
3
4

Figure 8-1: Test Station Components

8-2
Revision D
MOCON, Inc.