Chamber Specifications; Gas Supply Requirements; Sample Conditioning Capabilities - Ametek mocon AQUATRAN 3/40 Operator's Manual

Water vapor transmission rate system
Table of Contents

Advertisement

Specifications

Chamber Specifications

Uncrated

Gas Supply Requirements

Gas Composition
Supply Pressure, Nominal
Supply Pressure, Maximum
Carrier Gas
Supply Pressure, Minimum
Total Flow Rate, Nominal, Test
Active
Gas Composition
Supply Pressure, Nominal
Test Gas
Supply Pressure, Maximum
Supply Pressure, Minimum
Flow Rate, Maximum
Note: The low end of the Test Gas RH range requires a compressed air supply with the dew point of less
than -22 °C and an ambient humidity of less than 50% RH.

Sample Conditioning Capabilities

Temperature Control
Humidity Control, Chamber
Testing with 100%RH
10-2
Height
Width
17 inches
25.5 inches
(43.2 cm)
(64.8 cm)
Table 10.3: Chamber Specifications
Table 10.4: Gas Supply Specifications
Range
10 °C to 50 °C
Repeatability
± 0.2 °C at a single point within the chamber.
Accuracy
+/-0.5°C at a single point within the chamber.
10 - 23 °C
23 - 41 °C
Range
41 - 50 °C
10-50°C
Accuracy
± 5% RH
Use special 100%RH Test Cartridge for 100%RH generation
Table 10.5: Sample Conditioning Capabilities
Revision D
AQUATRAN Model 3/40 Operator's Manual
Depth
Volume
12 inches
3 cubic feet
(30.5 cm)
(85 Liters)
Nitrogen
25 PSI, (1.72 Bar), (172 kPa)
30 PSI, (2.07 Bar), (207 kPa)
23 PSI, (1.59 Bar), 159 kPa)
130 cc/min (10cc/min/cell); 490cc/min/cell)
Dry Compressed Air
85 PSI, (5.86 Bar), (586 kPa)
95 PSI, (6.55 Bar)
72 PSI, (5 Bar)
0.88 CFM, (25 LPM)
Low end 30%; High end 75 to 90% RH
30% to 90% RH
Low end 30%; High end 70 to 90% RH
See Chamber RH Working Range
Chart below for details
ISO 8753-1:2010 [1:3:1]
MOCON, Inc.