AQUATRAN Model 3/40 Operator's Manual

Chapter 6: Testing Packages

This chapter contains information on how to test packages. Suggestions on how to maximize the accuracy
of your results and procedures describing how prepare for and perform a package test are discussed.
Read this chapter to learn about:

Testing Suggestions

Package Testing Methods
Package Mounting Methods
Rigid Package Test Cartridge Overview
Flexible Package Test Cartridge Overview
Using a Package Adapter Cartridge
Testing with Chamber Generated RH or with 100%RH (Different Cartridges)
Package Purging Overview
Package Conditioning Overview
Setting Up and Starting a Package Test
Monitoring and Controlling a Package Test
Testing Suggestions
The following information will assist you in testing packages with your instrument.
Test duplicate samples whenever possible.
Use good samples. Poor sealing of packages to the test fixture and pin holes can contribute
to false readings.
Use caution when recording data for a new or unknown specimen. Some materials approach
equilibrium slowly. When in doubt repeat the test.
Turning off the power and gas supplies to instrument is not recommended. If the instrument
is shutdown you will need to wait for it to outgas before testing can resume.
MOCON, Inc.
Revision D
Testing Packages
6-1