System Shutdown for Storage or Repair

When shutting down the instrument for relocation, long-term storage, or if the instrument is to be sent for
factory repairs, follow these instructions:
1.
End all active tests and verify that the sensor is set to Bypass.
2.
Unclamp and remove any Test Cartridges in the chamber.
3.
Set the Chamber Temperature Set Point to the approximate ambient temperature and allow
the instrument to come to ambient temperature.
CAUTION:
Remove the water from the Humidifiers see "Drying Out the Humidifier" in this chapter.
4.
Turn off power to the instrument.
5.
Disconnect the gas lines at the rear of the module and use a brass fitting plug (025-382) to
seal the Carrier gas connection fitting on the rear of the instrument.
8-6
The instrument's testing temperature must be at ambient temperature
with the gases flowing before turning the instrument off for a long
period of time. If the temperature is not at ambient, condensation may
form and damage to the instrument is likely.
AQUATRAN Model 3/40 Operator's Manual
Revision D
MOCON, Inc.