Specifications
Transmission Rate Range and Repeatability Capabilities
Testing Range
2.5 to 100 gm/(m
0.161 to 6.45 gm/(100
0.0125 to 0.50 gm/(pkg-day)
0.05 to 2.5 gm/(m
0.0032 to 0.161 gm/(100
0.00025 to 0.0125 gm/(pkg-day)
0.05 to 10 gm/(m
0.0032 to 0.645 gm/(100
0.00025 to 0.05 gm/(pkg-day)
Notes:
1.
The ReZero Frequency must be set to one.
2.
When testing samples in both cells the transmission rates must be within a factor of 5.
3.
Calibration will be performed at the flow rate the samples were tested at.
4.
Testing performed at 37.8 Degrees Centigrade.
5.
The Repeatability specifications assume the use of a 100% RH Test Gas.
6.
Repeatability is based on a film area of 50 cm
9-4
Carrier
Flow
2
-day)
100
-day)
2
sccm
-day)
2
100
2
-day)
sccm
2
-day)
10
-day)
2
sccm
Table 9-6: Range and Repeatability Capability
Minimum
Exam
Calibration
Time
Standard
Required
Certified Film
25
P/N 054-162
minutes
(yellow)
Certified Film
25
P/N 054-163
minutes
(Red)
Certified Film
45
P/N 054-164
minutes
(Blue)
2
.
Revision G
AQUATRAN 3/38 Operator's Manual
Repeatability
+/- 2.0% of Reading
+/-0.05 g/(m
• day)
2
+/- 2.0% of Reading or
+/-0.05 g/(m
2
• day)
whichever greater
MOCON, Inc.