Loading And Unloading A Test Cartridge; Figure 3-1: Installing A Test Cartridge - Ametek mocon AQUATRAN 3/40 Operator's Manual

Water vapor transmission rate system
Table of Contents

Advertisement

Preparing for a Test
5.
Adjust the tubing as required to ensure the filter is resting on the bottom of the reservoir.
6.
Make sure the reclaiming tube is put back into the headspace of the reservoir (tank).

Loading and Unloading a Test Cartridge

The AQUATRAN Model 3/40 uses a pneumatic system to clamp and unclamp a Test Cartridge. The Test
Cartridge must be manually loaded into and unloaded from the instrument.
To remove a Test Cartridge from the instrument, follow the instructions below (refer to Figure 3-1):
Open the outer and inner chamber doors.
Press the Clamp/Unclamp button in front of the Test Cartridge to be removed.
Grasp the Test Cartridge firmly and lift straight up.
To install a Test Cartridge in the instrument, follow the instructions below (refer to Figure 3-1):
1.
Open the outer and inner chamber doors.
2.
Orient the Test Cartridge so the Test Cartridge gas ports face the Test Cell gas ports.
3.
Hold the Test Cartridge firmly and lower it straight down over the clamp bar.
4.
Press the Clamp/Unclamp button in front of the Test Cartridge being installed.
3-8

Figure 3-1: Installing a Test Cartridge

Revision D
AQUATRAN Model 3/40 Operator's Manual