SECTION 1
1.1 Local Mode
In Local Mode, you will control BENCH iLS primarily using the front panel controls. However,
many SCPI commands are still available – only setting of setpoints and activating the output
are disallowed.
When the output is OFF, the set of available commands is:
VOLTage:PROTect <value>
VOLTage:PROTect?
CURRent:PROTect <value>
CURRent:PROTect?
POWer:PROTect <value>
POWer:PROTect?
OUTPut:AUTO <ON|OFF>
OUTPut:AUTO?
SYSTem:CONFiguration:SAVE
When the output is ON, the set of available commands is:
OUTPut?
OUTPut OFF
MEASure:VOLTage?
MEASure:CURRent?
Regardless of the output state, you can query Status and Error registers, as described in
Section 6.0.
OPERATING MODES
iLS600, iLS600-R and iLS1500 Series
Set voltage protection level (volts)
Query voltage protection level (volts)
Set current protection level (amps)
Query current protection level (amps)
Set power protection level (watts)
Query power protection level (watts)
Enable/disable the Auto-Start feature
Query present Auto-Start state
Save configuration data to non-volatile memory
Query output state
Deactivate output
Measure and return output voltage (volts)
Measure and return output current (amps)
Rev. 05.20.21_#1.0