SECTION 5

MISCELLANEOUS COMMANDS

BENCH iLS recognizes a number of miscellaneous SCPI commands.

5.0 *CLS

*CLS clears the Event registers of the Questionable Status, Operation Status, Temperature
Status, and Hardware Status Register structures, in addition to the Error/Event Queue.

5.1 *IDN?

*IDN? returns the identification string for this unit. The string will contain, separated by
commas: 1) the company name, Artesyn Power; 2) the model; 3) the serial numbers of the
main and interface boards; and 4) the software revision number for the main and interface
board processors. For example, a 100V 10A model may, in response to *IDN?, return:
"Artesyn Power,Bench 100-10 iLS,031418032001,1.00.0051/1.00.1361"

5.2 *RST

*RST executes a soft reset of the system. The output is deactivated, all error Conditions are
cleared (however, error Events are not), and the unit is returned to its Output Off state. The
operating mode is not changed.
5.3 SYSTem:PROMpt <ON|OFF>
If you are interacting directly with BENCH iLS and would like to have a prompt,
SYSTem:PROMpt enables (ON) or disables (OFF) the SCPI prompt. The prompt is OFF by
default. The prompt is a newline (0x0A) character.