APPENDIX
SYSTem:AOUTput:MODE <DISabled|PARallel|SERies>
function of the analog output of the power supply to facilitate series or parallel operation. In
parallel mode the analog output is proportional to the output current of the power supply, in
series mode the analog output is proportional to the output voltage of the power supply.
SYSTem:AOUTput:Mode?
of the power supply.
SYSTem:CAPability?
(DCPSUPPLY WITH MEASURE) is returned.
SYSTem:CONFiguration:SAVE
non-volatile memory. Contents consist of:
•
Analog current scaling select (3V/5V/10V)
•
Analog current calibration slopes
•
Analog voltage scaling select (3V/5V/10V)
•
Analog voltage calibration slopes
•
Auto-start (Enabled/Disabled)
•
Calibrated flag (Calibrated/Not Calibrated)
•
Control source (Local/Remote/Remote with Lock/Analog Voltage Control/Analog
Current Control/Analog Dual Control)
•
Conversion current user slope and offset
•
Conversion current factory slope and offset
•
Conversion voltage user slope and offset
•
Conversion voltage factory slope and offset
•
Model
•
Over-current protection level
•
Over-power protection level
•
Over-voltage protection level
•
Power setpoint
•
Remote sense (Enabled/Disabled)
— This command returns the current state of the analog output
— Display the unit capability string. For BENCH iLS, the string
— This command saves configuration parameters to
iLS600, iLS600-R and iLS1500 Series
— This command controls the
Rev. 05.20.21_#1.0
advancedenergy.com
50