Additional Scpi Commands; Additional Capabilities; Analog Dual Control Mode; Voltage Setpoint - Advanced Energy iLS600 Series User Manual

Intelligent laboratory power 600w/1500w
Table of Contents

Advertisement

SECTION 1
VOLTage?
CURRent?
POWer?
Voltage is returned in volts, current in amperes, and power in watts. The output state is
either ON or OFF.

1.5.5 Additional SCPI Commands

Regardless of the output state, you can query Status and Error registers, as described in
Section 6.0.

1.5.6 Additional Capabilities

When the output is OFF, you can switch operating modes, as described in Section 1.0.
When the output is ON, you will be able to execute only those commands which activate
and deactivate the output and measure the output voltage and current.

1.6 Analog Dual Control Mode

1.6.1 Voltage Setpoint

When the unit is in Analog Dual Control Mode, the Analog Voltage channel supplies the
Voltage Setpoint.
Analog Input Scaling:
You can select 3 volt, 5 volt, or 10 volt scaling for the analog voltage input. This is the value
which represents the maximum model-specific voltage. For example, if you have the 400V
2.5A model and have selected the 10V scaling, then 10 volts applied across the Analog
Voltage channel will result in a 400 volt setpoint. For the same scaling, 5 volts applied
across the Analog Voltage channel will result in a 200 volt setpoint.
If Analog Voltage Control Mode is entered via the front panel, you will be asked to select
which scaling you want. You may also select scaling with the SCPI command:
SYSTem:MODe:ASCale VOLTage,<3|5|10>
OPERATING MODES
iLS600, iLS600-R and iLS1500 Series
Query voltage setpoint (volts)
Query current setpoint (amps)
Query power setpoint (watts)
Rev. 05.20.21_#1.0