SECTION 6
SCPI Access
Access to the Error/Event Queue is via the following commands:
SYSTem:ERRor[:NEXT]?
SYSTem:ERRor:COUNt?
SYSTem:ERRor:CLEar

6.0.15 Error Condition Register

Associated with the Error/Event Queue is a 16-bit Error Condition Register. It contains the
present state of a number of sensed error conditions.
Error Conditions
The Error Conditions recognized by BENCH iLS, identified by their error code are listed
below. If multiple error conditions are present, the register will contain the sum of the
individual error values.
1
2
4
8
16
32
64
128
256
512
1024
2048
32,768
STATUS AND ERRORS
Over-current condition
Over-voltage condition
Over-power condition
Output board over-temperature condition
Fan stall condition
Output (loss of regulation) error condition
12V bias error condition
3.3V bias error condition
Primary (input) board temperature error condition
PFC failure condition
Watchdog condition
Self-test error condition
PFC failure pending condition
iLS600, iLS600-R and iLS1500 Series
Return the next queue entry
Return the present number of queue entries
Clear the queue
Rev. 05.20.21_#1.0