Watchdog Error; Self-Test Error; Output Error; Scpi Access - Advanced Energy iLS600 Series User Manual

Intelligent laboratory power 600w/1500w
Table of Contents

Advertisement

SECTION 6

6.0.5 Watchdog Error

A watchdog timeout has occurred and the processor has restarted itself.

6.0.6 Self-Test Error

A self-test error has occurred. Self-test is run automatically at power-up and can
additionally be commanded remotely.

6.0.7 Output Error

Regulation of the output has been lost.

6.0.8 SCPI Access

Access to the Questionable Status Register is via the following commands:
STATus:QUEStionable[:EVENt]?
STATus:QUEStionable:CONDition?
STATus:QUEStionable:ENABle <value> Set Enable register
STATus:QUEStionable:ENABle?

6.0.9 Questionable Status Fan-Out Registers

In addition, there are two fan-out structures feeding the Temperature Error and
Hardware Error inputs.
STATUS AND ERRORS
iLS600, iLS600-R and iLS1500 Series
Rtn Event register
Rtn Condition register
Rtn Enable register
Rev. 05.20.21_#1.0
advancedenergy.com
23