APPENDIX
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]?
present Voltage setpoint, in volts.
[SOURce:]VOLTage:PROTectection[:LEVel] <value>
voltage protection threshold to value, which is specified in volts. Note: If storage of this
threshold to non-volatile memory is desired, SYSTem:CONFiguration:SAVE must also be
executed.
[SOURce:]VOLTage:PROTectection[:LEVel]?
voltage protection threshold, in volts.
STATus:OPERation:CONDition?
Operation Status Condition Register. If multiple bits are set, the result is the sum of the set
bit values.
1
2
4
8
16
32
64
128
256
512
1024
2048
4096
8192
16384
— This command returns the present contents of the
Calibrating (not used)
Setting (not used)
Ranging (not used)
Sweeping (not used)
Measuring
Waiting for Trigger (not used)
Waiting for Armed (not used)
Correcting (not used)
Output Activated
Constant Voltage mode
Constant Current mode
Constant Power mode
Not used
Instrument Summary Bit (not used)
Program running (not used)
iLS600, iLS600-R and iLS1500 Series
— This command returns the
— This command sets the over-
— This command returns the present over-
Rev. 05.20.21_#1.0
advancedenergy.com
46