APPENDIX
64
Summary Bit (RQS)
128
Operation Status
*TST?
— This command processes the Self-Test command, executing the unit selftest and
returning the results. This command is identical to the command "TEST:SELFtest[:EXECUT]"
The returned results are:
0
No self-test error
1
Output Voltage/Current ADC failed
*WAI
— Because no concurrent command execution is allowed, the command performs no
function.
CALibration:CALCulate:CURRent:PARameters
<setpoint1>,<measured1>,<setpoint2>,<measured2>
parameters provided to compute slope/offset calibration parameters for current control.
CALibration:CALCulate:VOLTage:PARameters
<setpoint1>,<measured1>,<setpoint2>,<measured2>
parameters provided to compute slope/offset calibration parameters for voltage control.
MEASure[:SCALar]:CURRent[:DC]?
output current. Output current is measured and buffered every 100 msecs. The returned
value is in amperes.
MEASure[:SCALar]:VOLTage[:DC]?
output voltage. Output voltage is measured and buffered every 100 msecs. The returned
value is in volts.
OUTPut:AUTOstart <ON|OFF>
either ON or OFF.
Note: This flag determines whether or not the unit automatically activates its output at power-up. This
command sets or clears the flag in volatile memory. For the modified flag to be valid after cycling primary
power, it must first be additionally stored to non-volatile memory, which is done with the
SYSTem:CONFiguration:SAVE command.
— This command returns the most recently measured
— This command returns the most recently measured
— This command sets the configuration auto-start flag
Rev. 05.20.21_#1.0
iLS600, iLS600-R and iLS1500 Series
— This command uses the four
— This command uses the four
advancedenergy.com
43