APPENDIX
*OPC?
— This command responds with an indication of whether the current operation is
complete. Because this unit does not support concurrent command execution, this
command will always return the value 1.
*RST
— This command executes a soft reset of the system. The output is deactivated, all
error Conditions are cleared (however, error Events are not), and the unit is returned to its
Output Off state. The operating mode (e.g., LOC or REM) is not changed.
*SRE <value>
— This command sets the Service Request Enable Register to the specified
value. The value is specified in decimal form with the following value assignments:
1
Reserved (not used)
2
Reserved (not used)
4
Error/Event Queue
8
Questionable Status
16
Message Available (not used)
32
Standard Event Status
64
Not user settable
128
Operation Status
The command *SRE 0 clears this register.
*SRE?
— This command returns the present contents of the Service Request Enable
Register. The contents are described in the description of *SRE. If multiple bits are set, the
result is the sum of the set bit values.
*STB?
— This command returns the present contents of the Status Byte Register. If multiple
bits are set, the result is the sum of the set bit values.
1
Reserved (not used)
2
Reserved (not used)
4
Error/Event Queue
8
Questionable Status
16
Message Available (not used)
32
Standard Event Status
iLS600, iLS600-R and iLS1500 Series
Rev. 05.20.21_#1.0
advancedenergy.com
42