Current Setpoint; Activating/Deactivating The Output; Remote Queries - Advanced Energy iLS600 Series User Manual

Intelligent laboratory power 600w/1500w
Table of Contents

Advertisement

SECTION 1

1.6.2 Current Setpoint

When the unit is in Analog Dual Control Mode, the Analog Current channel supplies the
Current Setpoint.
Analog Input Scaling:
You can select 3 volt, 5 volt, or 10 volt scaling for the analog current input. This is the value
which represents the maximum model-specific current. For example, if you have the 400V
2.5A model and have selected the 10V scaling, then 10 volts applied across the Analog
Current channel will result in a 2.5 amp setpoint. For the same scaling, 5 volts applied
across the Analog Current channel will result in a 0.750 amp setpoint.
If Analog Current Control Mode is entered via the front panel, you will be asked to select
which scaling you want. You may also select scaling with the SCPI command:
SYSTem:MODe:ASCale CURR <3|5|10>

1.6.3 Activating/deactivating the output

You can activate or deactivate the output either from the front panel Enable switch or by the
SCPI command:
OUTPut <ON|OFF> Activate/deactivate output

1.6.4 Remote Queries

When the output is ON, you may query the output current and voltage, as well as the output
state and voltage and current setpoints using the commands:
MEASure:VOLTage?
MEASure:CURRent?
OUTPut?
VOLTage?
CURRent?
POWer?
Voltage is returned in volts, current in amperes, and power in watts. The output state is
either ON or OFF.
OPERATING MODES
iLS600, iLS600-R and iLS1500 Series
Measure and return output voltage (volts)
Measure and return output current (amps)
Query output state
Query voltage setpoint (volts)
Query current setpoint (amps)
Query power setpoint (watts)
Rev. 05.20.21_#1.0
advancedenergy.com
14

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ils600-r seriesIls1500 series

Table of Contents