Power Meter Settings

Please find the sections below to find the general power meter
measurement settings.
Setting the Power Meter Top and Bottom End-Points
Setting the end points (Disk Range) close to expected
measurement value changes the sensitivity of the Power Meter
scale resolution. This is an advantage in viewing small changes in
power. However, this will not affect the overall range of the sensor.
This is only available for Meter and Chart display mode. For Trace
display, please press Meas > Peak Setup > Trace Setup to set the X
and Y scale.
Press Meas > Meas Disp > Dis Range to access the end-points
1
menu.
Press Top and enter the maximum scale value using the numeric
2
keypad. Then press dBm to complete the setup.
Press Bottom and enter the minimum scale value desired using the
3
numeric keypad. Then press dBm to complete the setup.
Setting Power Meter Resolution
There are four levels of power meter resolution. Higher resolutions
provide more accuracy but slow the measurement speed. The
resolution is only available for U2000 series power meter and
free-run trigger mode in U2020 series power meter.
Press Meas > Meas Disp > Resolution. Select 1,2,3 or 4. The four
options represent different resolution as follow:
• [1] = 1
• [2] = 0.1
• [3] = 0.01
• [4] = 0.001
Making a Power Measurement with USB Power Sensor
Functions and Measurements
81

Advertisement

Table of Contents
loading