Keysight Technologies N9321C User Manual page 161

Spectrum analyzer
Table of Contents

Advertisement

Key Reference
Peak Search
To Center
This is a hot key of marker to center.
Key access: Peak Search > Pk-Pk Search
Continuous Pk
When Continuous Pk is on, the marker will remain on the peak signal
On Off
even if the frequency of that signal changes. The marker will remain
on that signal as long as the amplitude of the signal does not
change by more than 3 dB from one sweep to another.
If the signal is lost, the marker will attempt to find it again and
maintain on the peak. If there are other signals on screen near the
same amplitude, one of them may be found instead. Signals near 0
Hz cannot be maintained effectively.
This function is intended to maintain the marker on signals with a
frequency that is changing, and an amplitude that is not changing.
Key access: Peak Search > Continuous PK
Peak Table
Peak Table
Displays a list of up to ten signal peaks that is updated
at the end of each sweep. The peaks can be sorted in order by
descending amplitude or by ascending frequency. Peaks above or
below the display line can be excluded from the table. The peak
table function works with trace 1 only.
Exprt Table to CSV
Press this softkey to save the current peak table
to a CSV file for your data analysis on PC.
Key access: Peak Search > More