Power Meter - Keysight Technologies N9321C User Manual

Spectrum analyzer
Table of Contents

Advertisement

Power Meter

Press Mode > Power Meter to access the optional power meter
functions. When this mode is selected, the power meter related
softkey menu will be displayed and many of the instrument
hardkeys will be disabled as they are not necessary for this mode.
Mode
toggles the measurement between AVG and Peak. When it is set to
AVG, the result shown in screen is the power sensor's
measurement result. When it is set to Peak, the result shown in
screen is the power sensor's gate measurement result.
Key access: Mode > Power Meter
Zeroing
Accesses the associated Zeroing submenu.
Key access: Mode > Power Meter
Zeroing
Zero Type
Calibration Toggles the zeroing type of the Power Sensor between
Makes a Zero operation of the power sensor.
The U2000 Series USB Power Sensors have their
sensor calibration tables stored in non- volatile
memory which means that the reference calibration
factor is automatically loaded upon power-up.
Key access: Mode > Power Meter > Zero/Cal
Toggles the zeroing type of the Power Sensor between
internal and external. When this type is set to EXT,
pressing Zero will bring up a message which indicates
the external signal need be pulled out. This setting is
not affected by the preset in Power Meter menu.
Key access: Mode > Power Meter > Zero/Cal
internal and external. When this type is set to EXT,
pressing Zero will bring up a message which indicates
the external signal need be pulled out. This setting is
not affected by the preset in Power Meter menu.
Key access: Mode > Power Meter > Zero/Cal
Key Reference
MODE
137