Keysight Technologies N9321C User Manual page 71

Spectrum analyzer
Table of Contents

Advertisement

4
Set the resolution bandwidth to 10 kHz:
Press BW > 10 > KHz
Excessive signal input may damage the DUT. Do not exceed the
C A U T I O N
maximum power that the device under test can tolerate.
5
Turn on the tracking generator and, if necessary, set the output
power to -10 dBm
Press Mode > Tracking Generator > Amplitude (On) > -10 > dBm
6
Connect the test cable (but not the DUT) from the tracking
generator output to the analyzer input. Store the frequency
response into trace 3 and normalize:
Press Meas > Normalize > Store Ref (1
7
Reconnect the DUT to the analyzer. Note that the units of the
reference level have changed to dB, indicating that this is now a
relative measurement.
8
To change the normalized reference position:
Press Meas > Normalize > Norm Ref Posn > 8
9
Place the reference marker at the specified cutoff frequency:
Press Marker > Mode > Delta Pair (Ref) > 370 > MHz
10
Place the second marker at 37 MHz:
Press Delta Pair (Delta) > 37 > MHz
11
In this example, the attenuation over this frequency range is -17.98
dB/octave (one octave above the cutoff frequency).
Measuring Stop Band Attenuation of a Lowpass Filter
:
Functions and Measurements
4) > Normalize (On)
63