ratio by using the arrow keys, knob, or numeric keypad. When
VBW/RBW (Auto) is selected, the ratio is determined automatically
by the rules in
Figure 5VBW/RBW Ratio Auto Rules
Key access: BW > VBW/RBW
Average Type
Chooses one of the following averaging scales: log power (video),
power (RMS), or voltage averaging. When trace average is on, the
average type is shown on the left side of the display. Selecting an
average type manually causes the analyzer to use that type without
regard to other analyzer settings.
Key access: BW > Average Type
Figure 5
.
Yes
Is Detector peak
or Negative peak?
No
No
Is Detector
Sample?
Yes
No
Is Avg Type
Ratio=1
Power?
Yes
Is Trace
Yes
Ratio=10
Average On?
No
Ratio=1
Key Reference
Yes
Is Trace
Ratio=10
Average On?
No
Ratio=1
Yes
Is Avg Type
Ratio=10
Power?
No
Ratio=1
BW
109