If the signal peak is not present on the display, increase the
frequency span out to 20 MHz, turn signal tracking on, decrease
the span back to 2 MHz and turn signal tracking off:
Press Span > 20 > MHz
Press Peak Search
Press Span > 2 > MHz
Figure 3-5 Unresolved Signals of Equal Amplitude
4
Change the resolution bandwidth (RBW) to 100 kHz so that the
RBW setting is less than or equal to the frequency separation of the
two signals:
Press BW > 100 > kHz
Notice that the peak of the signal has become flattened indicating
that two signals are present.
5
Decrease the video bandwidth to 3 kHz:
Press BW > Video BW > 3 > kHz
Two signals are now visible as shown in
front-panel knob or arrow keys to further reduce the resolution
bandwidth and better resolve the signals.
Functions and Measurements
Measuring Multiple Signals
Figure 3-6
. Use the
39