Keysight N9000A CXA Signal Analyzer
Option T03 and T06
Tracking Generator
Notice: This document contains references to Agilent.
Please note that Agilent's Test and Measurement
business has become Keysight Technologies. For
more information, go to www.keysight.com.
Installation Note
Part Number: N9000-90032
Printed in USA
July 2014
*N9000-90032*
N9000-90032

Advertisement

loading
Need help?

Need help?

Do you have a question about the N9000A and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Keysight Technologies N9000A

  • Page 1 Keysight N9000A CXA Signal Analyzer Option T03 and T06 Tracking Generator Notice: This document contains references to Agilent. Please note that Agilent’s Test and Measurement business has become Keysight Technologies. For more information, go to www.keysight.com. Installation Note Part Number: N9000-90032...
  • Page 2 Notice. The information contained in this document is subject to change without notice. t Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
  • Page 3 Overview This kit contains all of the instructions required to install Option T03 and T06 tracking generator on an Keysight N9000A CXA Signal Analyzer. The procedure outlined in this document will guide you through the following steps, which will need to be followed closely, and performed in the order defined: 1.
  • Page 4: Installation Procedure

    Option T03, T06 - Tracking Generator Installation Kit Installation Procedure This installation note is part of an upgrade kit that provides all of the parts and instructions needed to install the tracking generator board assembly in an Agilent N9000A CXA Signal Analyzer. Contents Quantity...
  • Page 5 Option T03, T06 - Tracking Generator Installation Kit 1. Pre-Upgrade Status Check Before performing this upgrade you will need to know the status of the instrument being upgraded in a couple of areas. Pre-Existing Issues and Error Messages Power the instrument on and verify that the instrument will boot up properly. This would include verifying that there are no error messages displayed in the Status Bar of the display.
  • Page 6: Hardware Installation

    Option T03, T06 - Tracking Generator Installation Kit 2. Hardware Installation CAUTION Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic assemblies should be performed at a static-safe workstation. Refer to the documentation that pertains to your instrument for information about static-safe workstations and ordering static-safe accessories.
  • Page 7 Option T03, T06 - Tracking Generator Installation Kit b. To Remove the reference bracket (2) use the T-10 driver to remove the six screws (5) (including two at the rear panel) attaching the reference bracket to the instrument and the four screws (4) attaching the top brace to the reference bracket.
  • Page 8 Option T03, T06 - Tracking Generator Installation Kit 4. Refer to Figure 3 to remove the front frame assembly. Figure 3 Front Frame Assembly Removal a. Remove the overlay (2) on the front frame assembly. b. Remove the eight screws (1), four on each side, to detach the front frame assembly from the chassis.
  • Page 9 Option T03, T06 - Tracking Generator Installation Kit c. Pull the front frame assembly carefully away from the chassis. Refer to Figure 4 to remove the ribbon cable W1 from the mother board. Figure 4 Front Panel Cable Installation Note N9000-90032...
  • Page 10 Option T03, T06 - Tracking Generator Installation Kit Install the Tracking Generator Assembly 1. Locate the tracking generator assembly in the retrofit kit. 2. Refer to Figure 5 to install the assembly (3) into slot 4 in the chassis and press down to plug it into the motherboard.
  • Page 11 Option T03, T06 - Tracking Generator Installation Kit Install the Cable Assembly 1. Locate the cable assembly(N9000-20109) in the retrofit kit. 2. Refer to , attach the semi-rigid cable to RF input connector assembly and tighten the SMA Figure 7 connector with 10 in-lbs.
  • Page 12 Option T03, T06 - Tracking Generator Installation Kit 4. Locate the cable assemblies N9000-20104, N9000-20105, N9000-20106, and N9000-20107 in the retrofit kit. 5. Refer to , install N9000-20105 (2), N9000-20106 (3), N9000-20104 (4), and N9000-20107 Figure 8 (5) onto the corresponding instrument assemblies. 6.
  • Page 13 Option T03, T06 - Tracking Generator Installation Kit 3. Update the Instrument Software 1. Load the latest instrument software is required to assure all FPGAs and drivers located on both the newly installed hardware and on the case instrument are synchronized. Therefore, even if the instrument contains the latest revision of software, you must re-install the software to assure proper operation.
  • Page 14 Option T03, T06 - Tracking Generator Installation Kit 6. Installation Verification Power the instrument on and verify that the instrument will not only boot up, but that the signal analyzer application software will load properly. This would include verifying that there are no error messages displayed in the Status Bar of the application window.