 Reference or digital wave form
 Cursor
 Trigger control
 FFT
Operation steps:
1. Push the Display button.
2. Select XY Mode in the bottom menu. Select Enable as ON in the right menu.
3.
To make the XY view full screen, select Full Screen as ON in the right menu.
Counter
It is a 6-digit single-channel counter. The counter can only measure the frequency of
the triggering channel. The frequency range is from 2Hz to the full bandwidth. Only if
the measured channel is in Edge mode of Single trigger type, the counter can be
enabled. The counter is displayed at the bottom of the screen.
Operation steps:
1. Push Trigger Menu button, set the trigger type to Single, set the trigger mode to
Edge, select the signal source.
2. Push the Display button.
3. Select Counter as ON or OFF in the bottom menu.

How to Save and Recall a Waveform

Push the Save button, you can save the waveforms, configures, screen images, record
or clone the waveform.
The description of the Save Function Menu is shown as the following table:
Function Menu
Type
When the type is Wave, the menu shows as following:
Type
Wave
4.Advanced User Guidebook
Setting
Description
Wave
Choose the saving type.
Configure
About the Record type, see "How to
Image
Record/Playback Waveforms" on P73.
Record
About the Clone type, see "How to Clone
Clone
and Recall a waveform" on P77.
Format
For internal storage, only BIN can be
selected. For external storage, the format
(Right
can be BIN, TXT or CSV.
menu)
66