Owon XDS3204E User Manual page 51

Digital storage oscilloscopes
Table of Contents

Advertisement

when
Mode
Holdoff
5.Runt Trigger
Trigger pulses that pass through one trigger level but fail to pass through the other
trigger level. Shown as below figure,
In Runt Trigger mode, the trigger setting information is displayed on bottom right of
the screen, for example,
trigger source is CH1, polarity is positive, 0.00mV is the differential between up level
and low level threshold.
Runt Trigger menu list:
MENU
SETTING
4.Advanced User Guidebook
Polarity
Choose the polarity
Select pulse width condition and adjust the M knob
or click
button or click
which digit to be set.
Auto
Acquire waveform even no trigger occurred
Normal
Acquire waveform when trigger occurred
Single
When trigger occurs, acquire one waveform then stop
100 ns - 10 s, adjust M knob or click
Holdoff
interval before another trigger occur, press
panel button or click
choose which digit to be set.
Reset
Set Holdoff time as 100 ns
Runt Trigger
INSTRUCTION
to set time, press
,indicates that trigger type is runt,
44
to move cursor to choose
to set time
to move cursor to
panel