Using the Detector
Special Settings of the Detector

Configuration Settings

Figure 28
Detector Configuration Settings
If the flow cell temperature is critical for your chromatography or your environment is
N O T E
stable, you may set the Temperature Control to off. This will lower the optical unit and flow
cell temperature by some degree.
72
• Temperature Control: the optical unit is kept on constant
temperature (some degrees above ambient) and improves
the baseline stability in unstable environments. See also
note below.
• UV lamp tag: Automatic mode for Agilent lamps with
RFIDtags. If no RFID tag lamp is used, detector icon will
become gray (lamp tag not ready) and analysis is disabled.
The specifications are based on the lamp with RFID tag.
Use UV lamp anyway - here you can select non-RFID tag
lamps, such as of type VWD lor DAD (have different
heating mechanismn).
The correct selection is important for optimal performance
and lifetime.
• Cell tag: for Agilent flow cells with RFID tags. If no RFID
tag cell is used, detector icon will become gray (cell tag not
ready) and analysis is disabled.
• Help: online help.
The detector status shows "Cell-tag" in yellow in case the
flow cell with the RFID tag is not inserted. The detector icon is
gray and the system is not ready.
Agilent 1200 Infinity Series VWD - User Manual