Requirements and Conditions

What You Will Need
The table below lists the items you need to have for the set up of the
analysis. Some of these are optional (not required for the basic system).
Table 8
Agilent 1200
Infinity Series
system
Column:
Standard:
Conditions
A single injection of the isocratic test standard is made under the
conditions given in
Table 9
Flow
Stoptime
Solvent
Temperature
Wavelength
Injection Volume
Column Temperature (optional):
Agilent 1200 Infinity Series VWD - User Manual
What you will need
Pump (plus degassing)
Autosampler
Detector, standard flow cell installed
Degasser (optional)
Column Compartment (optional)
Agilent ChemStation or
Instant Pilot G4208, optional for basic operation.
System should be correctly set up for LAN communication with the Agilent
ChemStation
Zorbax Eclipse XDB-C18, 4.6 x 150 mm, 5 µm (993967-902) or an equivalent column
Agilent isocratic checkout sample (01080-68704)
Table 9
Conditions
on page 59:
1.5 mL/min
8 min
100% (30% water/70% Acetonitrile)
Ambient
sample 254 nm
1 µL
25 °C or ambient
Using the Detector
Setting up an Analysis
4
59