Compensation Sensor Open; Compensation Sensor Short - Agilent Technologies 1200 Infinity Series User Manual

Hide thumbs Also See for 1200 Infinity Series:
Table of Contents

Advertisement

7
Error Information
General Error Messages

Compensation Sensor Open

Error ID: 0081
The ambient- compensation sensor (NTC) on the main board in the module
has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the
main board is dependent on ambient temperature. The change in
resistance is used by the leak circuit to compensate for ambient
temperature changes. If the resistance across the sensor increases above
the upper limit, the error message is generated.
Probable cause
1

Compensation Sensor Short

Error ID: 0080
The ambient- compensation sensor (NTC) on the main board in the module
has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the
main board is dependent on ambient temperature. The change in
resistance is used by the leak circuit to compensate for ambient
temperature changes. If the resistance across the sensor falls below the
lower limit, the error message is generated.
Probable cause
1
102
Defective main board.
Defective main board.
Suggested actions
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Agilent 1200 Infinity Series VWD - User Manual