Remote Timeout

Error ID: 0070
A not- ready condition is still present on the remote input. When an
analysis is started, the system expects all not- ready conditions (for
example, a not- ready condition during detector balance) to switch to run
conditions within one minute of starting the analysis. If a not- ready
condition is still present on the remote line after one minute the error
message is generated.
Probable cause
1
2
3
Agilent 1200 Infinity Series VWD - User Manual
Leak detected in another module with a
CAN connection to the system.
Leak detected in an external instrument
with a remote connection to the system.
Shut-down in an external instrument with a
remote connection to the system.
The degasser failed to generate sufficient
vacuum for solvent degassing.
Not-ready condition in one of the
instruments connected to the remote line.
Defective remote cable.
Defective components in the instrument
showing the not-ready condition.
Error Information
General Error Messages
Suggested actions
Fix the leak in the external instrument before
restarting the module.
Fix the leak in the external instrument before
restarting the module.
Check external instruments for a shut-down
condition.
Check the vacuum degasser for an error
condition. Refer to the Service Manual for the
degasser or the 1260 pump that has the
degasser built-in.
Suggested actions
Ensure the instrument showing the not-ready
condition is installed correctly, and is set up
correctly for analysis.
Exchange the remote cable.
Check the instrument for defects (refer to the
instrument's documentation).