Error Information
Detector Error Messages

Heater Power At Limit

Error ID: 1074
The available power of the heater reached either the upper or lower limit.
This event is sent only once per run. The parameter determines which
limit has been hit:
0 means upper power limit hit (excessive ambient temperature drop).
1 means lower power limit hit (excessive ambient temperature increase).
Probable cause
1
No Run Data Available In Device
In a very rare case the capacity of the CompactFlash Card is not
sufficient. This could happen for example when the interrupt of LAN
communication takes longer and the detector uses special settings (e.g full
data rate at 80 Hz plus full spectra plus all signals) during data buffering.
Probable cause
1
112
Excessive ambient temperature change.
CompactFlash Card is full.
Suggested actions
Wait until temperature control equilibrates.
Suggested actions
• Correct communication problem.
• Reduce data rate.
Agilent 1200 Infinity Series VWD - User Manual