Instrument Commands - Keysight M8121A User Manual

Arbitrary waveform generator
Table of Contents

Advertisement

6
Remote Programming

:INSTrument Commands

Parameter Suffix
Parameter Suffix
92
:INSTrument:COUPle:STATe[1|2][?]
Command
:INSTrument:COUPle:STATe[1|2] OFF|ON|0|1
Query
:INSTrument:COUPle:STATe[1|2]?
Parameters
OFF|ON|0|1
None
Description
This command switch coupling on/off. Coupled mode means both
channels are synchronized. If coupling is switched from on to off, the
values of the channel where coupling is switched on are taken. The
coupled values are:
:TRAC[1|2]:DWID
Examples
Command
:INST:COUP:STAT1 ON
Query
:INST:COUP:STAT1?
:INSTrument:SLOT[:NUMBer]?
Query
:INSTrument:SLOT[:NUMBer]?
Parameters
None
None
Description
This query returns the instrument's slot number in its AXIe frame.
Examples
Query
:INST:SLOT?
Keysight M8121A Arbitrary Waveform Generator User Guide

Advertisement

Table of Contents
loading

Table of Contents