Calibrate Commands - Keysight M8121A User Manual

Arbitrary waveform generator
Table of Contents

Advertisement

6
Remote Programming

:CaLibrate Commands

Parameter Suffix
Parameter Suffix
132
The :CALibrate commands are used to control the latency from the
M8131A digitizer to the M8121A AWG. This set of commands should be
used only when controlling both instruments. The end effect is that the
delay from the analog input of the digitizer to the analog output of the
AWG is a fixed latency for a given setup. The calibration procedure must
be run upon a change in conditions, such as a disruption to the data flow
over ODI, change in cables, or a change in clock source.
:CALibrate:SYNC:INPut:LEVel
Command
:CAL:SYNC:INP:LEV <level>
Query
:CAL:SYNC:INP:LEV?
Parameters
<level>|MINimum|MAXimum|DEFault
None
Description
Set or query the Sync input threshold in Volts.
The Sync Input is used with the M8131A digitizer to calibrate a
deterministic latency from analog input to analog output.
Examples
:CALibrate:SYNC:INPut:LEVel 0.4
:CALibrate:LATency:LCMP
Command
:CALibrate:LATency:LCMP <lcmp_value>
Parameters
<lcmp_value>
None
Description
Lowest common multiplier between the internal clocks of the M8131A and
M8121A. Use value of 5 when operating with a M8131A digitizer.
Examples
:CALibrate:LATency:LCMP 5
Keysight M8121A Arbitrary Waveform Generator User Guide