Power and Ventilation Requirements

Thermal Protection

Battery

Operating Environment

Keysight M8121A Arbitrary Waveform Generator User Guide
For power and ventilation requirements, refer to:
http://www.keysight.com/find/M9514A
documentation.
http://www.keysight.com/find/M9505A
documentation.
http://www.keysight.com/find/M9502A
documentation.
Overheating Detection
The instrument monitors its internal temperature. If the temperature
exceeds what the instrument can tolerate, the power supply is switched
off. The instrument will not turn on automatically if the temperature is
decreasing again.
Fan Failure
If a fan is broken or prevented from operating by a blockage the
temperature will increase. When the temperature exceeds what the
instrument can tolerate, the overheating detection switches off the
instrument for safety reasons. For reliability it is recommended to send
instruments with broken or defective fans immediately to Keysight Service
for repair.
The M8121A module does not have a battery.
For details on the operating environment for the M8121A module, refer to
the section
Operating Environment
Installation and Maintenance
for 14-slot chassis related
for 5-slot chassis related
for 2-slot chassis related
on page 139.