Recommended Test Equipment; Service Strategy; To Clean The Instrument - Keysight 1143A User's And Service Manual

Probe offset control and power module
Hide thumbs Also See for 1143A:
Table of Contents

Advertisement

Table 5

Recommended Test Equipment

The following table is a list of the test equipment required to service this
instrument. The table indicates the critical specification of the test
equipment and for which procedure the equipment is necessary. Equipment
other than the recommended model may be used if it satisfies the critical
specification listed in the table.

Recommended Test Equipment

Equipment
Required
Critical Specifications
Accuracy, ±0.1%
DVM
511 Ω, 0.5 W, 1%
Resistor
50 Ω, 10 W, 1%
Resistor
P = Performance Tests, A = Adjustments, T = Troubleshooting

Service Strategy

The 1143A Probe Offset Control and Power Module consists of simple power
supplies and op-amp variable current sources. Circuitry is simple and
components are available off the shelf, so the service strategy is
component-level repair. The troubleshooting, parts lists, theory of operation,
and schematics support this repair level.

To clean the instrument

Use mild soap and water to clean the instrument. Harsh soaps will damage
the water-based paint finish of the instrument.
Service
General Information
Recommended
Model/Part
Use
3457A
P,A,T
0757-0814
A
0811-3707
A
23