Inspection And Cleaning; Visual Inspection; Exterior Cleaning; Interior Cleaning - GE MAC 5000 Service Manual

Resting ecg analysis system
Hide thumbs Also See for MAC 5000:
Table of Contents

Advertisement

Inspection and Cleaning

Visual Inspection

Exterior Cleaning

Interior Cleaning

General

4-4
Maintenance: Inspection and Cleaning
Perform a visual inspection of all equipment and peripheral devices
daily. Turn off the unit and remove power before making an inspection or
cleaning the unit.
Check the case and display screen for cracks or other damage.
Regularly inspect all cords and cables for fraying or other damage.
Verify that all cords and connectors are securely seated.
Inspect keys and controls for proper operation.
Toggle keys should not stick in one position.
Knobs should rotate fully in both directions.
Clean the exterior surfaces monthly, or more frequently if needed.
1. Use a clean, soft cloth and a mild dish washing detergent diluted in
water.
2. Wring the excess water from the cloth. Do not drip water or any
liquid on the equipment, and avoid contact with open vents, plugs, or
connectors.
3. Dry the surfaces with a clean cloth or paper towel.
Check for dust buildup on the surfaces of the interior circuit boards,
components, and power supply. Use commercially available compressed
air to blow away the accumulated dust. Follow the manufacturers
directions.
MAC™ 5000 resting ECG analysis system
2024917-010
Revision B