Clean the Detector Exterior

WA RNING
WA RNING
88
Burn hazard. The burner assembly can be hot enough to cause
burns. Before touching, cool to a safe handling temperature
(< 40 °C).
Shock hazard. Before cleaning the detector, turn it off and unplug
its power cord.
Before cleaning the detector, shut it down, turn it off, and
unplug the detector power cord. Clean the detector with a damp
cloth using water. Do not spray liquids directly onto the
detector. Wipe dry with a clean, soft cloth. Do not allow cleaning
fluids to drip into the detector or GC because liquids can
damage the detector or GC electronics.
Do not use cleaning agents on the burner assembly that could
cause a hazard on the burner.