GC keyboard control

Identifying the detector

50
To access the method parameters for an SCD or NCD, press the
[Front Det], [Back Det], or [Aux Det #] (tandem FID-XCD
configuration only) key. See
To turn on the PMT voltage, press [Config], then press the key for
the detector ([Front Det], [Back Det], or [Aux Det #]). The PMT
voltage can be turned on or off. The PMT operates at a constant
voltage (800 V).
Table 5
lists the expected configurations for an XCD.
Table 5
Connections to the GC wiring harnesses
XCD Installation
Front XCD
Back XCD
Single front FID-XCD (front tandem
FID-XCD)
Single back FID-XCD (back tandem
FID-XCD)
Dual front FID-XCD and back FID-XCD
(Dual tandem FID-XCD)
Table
5.
Which detector is the XCD?
Front
Back
Front: Front FID
Aux detector 2: XCD
Back: Back FID
Aux detector 2: XCD
Front: Front FID
Aux detector 1: front XCD
Back: Back FID
Aux detector 2: back XCD