Configure Auto Flow Zero on the GC

SCD and NCD User Manual
Agilent recommends setting the GC to automatically zero flow
sensors to reduce drift. See the GC Operation Manual for
details.
On the GC keypad, press [Options].
1
Scroll to Calibration and press [Enter].
2
3
Scroll to select the appropriate detector (front, back, aux 2,
or aux 1) and press [Enter].
Scroll to Autoflow zero (H2 Lower) and press [On/Yes]. (To turn
4
off autozero, instead press [Off/No].
For SCD, repeat for Autoflow zero (H2 Upper).
5
4
Operation
57